• DocumentCode
    3533892
  • Title

    Comparison of three electrical capacitance tomography systems

  • Author

    Chen, Dixiang ; Yang, Wuqiang ; Deng, Xiang

  • Author_Institution
    Sch. of Mechatron. Eng. & Autom., Nat. Univ. of Defense Technol., Changsha, China
  • fYear
    2010
  • fDate
    1-2 July 2010
  • Firstpage
    57
  • Lastpage
    62
  • Abstract
    Electrical capacitance tomography (ECT) has been developed for two decades to visualize the internal behavior of industrial processes containing dielectric components. Three different ECT systems have been developed: (1) based on a charge/discharge capacitance measuring circuit, (2) based on an AC circuit and (3) based on an impedance analyzer. In this paper the operation principles, calibration methods and some performance of the three systems are compared. The experimental results prove that the AC-based ECT system is superior to the charge/discharge ECT system. Although the impedance analyzer based ECT system can provide the highest accuracy and the best image quality, it is bulky and expensive.
  • Keywords
    calibration; capacitance measurement; dielectric devices; electric impedance imaging; tomography; AC circuit; ECT system; charge-discharge capacitance measuring circuit; dielectric components; image quality; impedance analyzer; industrial processes; three electrical capacitance tomography systems; Calibration; Capacitance measurement; Charge measurement; Circuits; Current measurement; Dielectric measurements; Electrical capacitance tomography; Image analysis; Impedance measurement; Visualization; ECT; calibration; charge/discharge; impedance analyzer;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Imaging Systems and Techniques (IST), 2010 IEEE International Conference on
  • Conference_Location
    Thessaloniki
  • Print_ISBN
    978-1-4244-6492-0
  • Type

    conf

  • DOI
    10.1109/IST.2010.5548441
  • Filename
    5548441