DocumentCode :
3538451
Title :
The application of Empirical Mode Decomposition in elimination of ECG contamination from EMG signals
Author :
Naji, Mohsen ; Firoozabadi, Mohammad ; Kahrizi, Sedighe
Author_Institution :
Dezful Branch, Dept. of Biomed. Eng., Islamic Azad Univ., Dezful, Iran
fYear :
2011
fDate :
14-16 Dec. 2011
Firstpage :
77
Lastpage :
80
Abstract :
EMG signals collected from the trunk musculature are often contaminated with the heart muscle electrical activity (ECG). This paper introduces a novel method, the Empirical Mode Decomposition, for elimination of ECG contamination from EMG signals. The method is compared to a Butterworth high pass filtering. Results obtained from the analysis of generated and experimental EMG signals show that our method outperforms the high pass filtering for elimination of ECG contamination from trunk EMG signals.
Keywords :
deconvolution; electrocardiography; electromyography; medical signal processing; Butterworth high pass filtering comparison; ECG contamination elimination; EMG signals; empirical mode decomposition; heart muscle electrical activity; trunk musculature; Contamination; Electrocardiography; Electrodes; Electromyography; Filtering; Muscles; Noise reduction;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Engineering (ICBME), 2011 18th Iranian Conference of
Conference_Location :
Tehran
Print_ISBN :
978-1-4673-1004-8
Type :
conf
DOI :
10.1109/ICBME.2011.6168590
Filename :
6168590
Link To Document :
بازگشت