DocumentCode
3543641
Title
Experimental results from an electrostrictive multiplexed 2-D array
Author
Davidsen, Richard E. ; Smith, Stephen W.
Author_Institution
Dept. of Biomed. Eng., Duke Univ., Durham, NC, USA
Volume
2
fYear
1997
fDate
5-8 Oct 1997
Firstpage
1647
Abstract
A 2:1 multiplexed two-dimensional array has been developed which has a sparse element pattern designed for real time volumetric imaging and a second element pattern designed for conventional B-mode imaging. For volumetric imaging a small aperture was used to provide a wide transmit beam, allowing multiple beams to be received simultaneously. A larger aperture with a more narrow transmit beam was used for B-mode imaging to improve image quality when multiple receive beams were not required. The multiplexed transducer was fabricated using an electrostrictive relaxor ferroelectric material in which array elements were activated and deactivated by a DC bias field
Keywords
biomedical transducers; biomedical ultrasonics; electrostriction; ferroelectric devices; multiplexing; ultrasonic transducer arrays; 2:1 multiplexed two-dimensional array; DC bias field; array elements; conventional B-mode imaging; electrostrictive multiplexed 2-D array; electrostrictive relaxor ferroelectric material; image quality; medical diagnosis; multiple beams; multiplexed transducer; real time volumetric imaging; second element pattern; small aperture; sparse element pattern; volumetric imaging; wide transmit beam; Apertures; Biomedical engineering; Biomedical imaging; Electrostriction; Ferroelectric materials; Image quality; Piezoelectric materials; Piezoelectric polarization; Relaxor ferroelectrics; Ultrasonic imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 1997. Proceedings., 1997 IEEE
Conference_Location
Toronto, Ont.
ISSN
1051-0117
Print_ISBN
0-7803-4153-8
Type
conf
DOI
10.1109/ULTSYM.1997.663310