• DocumentCode
    3545268
  • Title

    Heart rate control by the computerized infusion of isoproterenol

  • Author

    Leor, R. ; Bobrovsky, BZ ; Eliash, S. ; Kaplinsky, E.

  • Author_Institution
    Dept. of Electr. Eng. Syst., Tel Aviv Univ., Israel
  • fYear
    1997
  • fDate
    7-10 Sep 1997
  • Firstpage
    517
  • Lastpage
    519
  • Abstract
    The authors developed a computerized system for the controlled acceleration of heart rate. Ten cannulated rats were connected to a monitoring system that sampled the ECG and arterial blood pressure. Isoproterenol was delivered via a computer controlled infusion pump that the authors have developed. A control software, based on a modified proportional integral controller, sensed the heart rate and stabilized it on a predetermined target heart rate. Stable heart rate values within ±9.3 BPM of target heart rate were achieved in all ten rats with average settling time of 6.4±4.3 minutes. The authors conclude that heart rate acceleration and stabilization by the computer controlled infusion of isoproterenol is feasible and will undoubtedly have practical future clinical applications
  • Keywords
    biocontrol; biomedical equipment; blood pressure measurement; cardiology; computerised control; electrocardiography; patient treatment; two-term control; 6.4 min; clinical applications; computerized infusion; control software; controlled heart rate acceleration; heart rate control; isoproterenol; modified proportional integral controller; rats; settling time; target heart rate; Acceleration; Application software; Arterial blood pressure; Computerized monitoring; Control systems; Electrocardiography; Heart rate; Pi control; Proportional control; Rats;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology 1997
  • Conference_Location
    Lund
  • ISSN
    0276-6547
  • Print_ISBN
    0-7803-4445-6
  • Type

    conf

  • DOI
    10.1109/CIC.1997.647948
  • Filename
    647948