• DocumentCode
    3545627
  • Title

    RF-link for cortical neuroprosthesis

  • Author

    Gerald, José A B ; Tavares, Gaoçalo ; Piedade, Moisés S. ; Varela, Elísio ; Ribeiro, Ricardo

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Inst. Superior Tecnico, Lisboa, Portugal
  • fYear
    2005
  • fDate
    23-26 May 2005
  • Firstpage
    2739
  • Abstract
    This paper describes an RF-link system suitable for intracortical neuroprosthesis. The wireless coupling minimizes the risk of infection and improves the patient health level. Data and power are transmitted by means of an RF low-coupling transformer. Binary FSK modulation is used in the forward link (from outside to implanted prosthesis) and BPSK is used in the backward link (implanted prosthesis to outside). Experimental results are presented, and important concerns regarding the magnetic coupling are addressed.
  • Keywords
    coupled circuits; frequency shift keying; high-frequency transformers; neurophysiology; phase shift keying; prosthetics; radio links; 1 MHz; 1 to 2 cm; 100 kbit/s; 35 percent; 50 mW; BPSK; RF low-coupling transformer; RF-link system; binary FSK modulation; implanted prosthesis; infection risk minimization; intracortical neuroprosthesis; magnetic coupling; wireless coupling; Clocks; Couplings; Demodulation; Electrodes; Frequency shift keying; Power amplifiers; Power engineering computing; Prosthetics; Radio frequency; Radiofrequency amplifiers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Circuits and Systems, 2005. ISCAS 2005. IEEE International Symposium on
  • Print_ISBN
    0-7803-8834-8
  • Type

    conf

  • DOI
    10.1109/ISCAS.2005.1465193
  • Filename
    1465193