DocumentCode
3545627
Title
RF-link for cortical neuroprosthesis
Author
Gerald, José A B ; Tavares, Gaoçalo ; Piedade, Moisés S. ; Varela, Elísio ; Ribeiro, Ricardo
Author_Institution
Dept. of Electr. & Comput. Eng., Inst. Superior Tecnico, Lisboa, Portugal
fYear
2005
fDate
23-26 May 2005
Firstpage
2739
Abstract
This paper describes an RF-link system suitable for intracortical neuroprosthesis. The wireless coupling minimizes the risk of infection and improves the patient health level. Data and power are transmitted by means of an RF low-coupling transformer. Binary FSK modulation is used in the forward link (from outside to implanted prosthesis) and BPSK is used in the backward link (implanted prosthesis to outside). Experimental results are presented, and important concerns regarding the magnetic coupling are addressed.
Keywords
coupled circuits; frequency shift keying; high-frequency transformers; neurophysiology; phase shift keying; prosthetics; radio links; 1 MHz; 1 to 2 cm; 100 kbit/s; 35 percent; 50 mW; BPSK; RF low-coupling transformer; RF-link system; binary FSK modulation; implanted prosthesis; infection risk minimization; intracortical neuroprosthesis; magnetic coupling; wireless coupling; Clocks; Couplings; Demodulation; Electrodes; Frequency shift keying; Power amplifiers; Power engineering computing; Prosthetics; Radio frequency; Radiofrequency amplifiers;
fLanguage
English
Publisher
ieee
Conference_Titel
Circuits and Systems, 2005. ISCAS 2005. IEEE International Symposium on
Print_ISBN
0-7803-8834-8
Type
conf
DOI
10.1109/ISCAS.2005.1465193
Filename
1465193
Link To Document