• DocumentCode
    3548137
  • Title

    Simultaneous topography imaging and molecular recognition with low crosstalk and high sensitivity

  • Author

    Haitao Yang ; Guangyong Li

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Univ. of Pittsburgh, Pittsburgh, PA, USA
  • fYear
    2013
  • fDate
    10-13 Nov. 2013
  • Firstpage
    136
  • Lastpage
    139
  • Abstract
    Atomic force microscope has been used to recognize biomolecules through a functionalize tip. In order to achieve simultaneous topography imaging and molecular recognition, existing method such as PicoTREC (commercialized by Agilent Inc.) extracts the topography image from the lower amplitude of the cantilever oscillation while extracts recognition image from the upper amplitude of the cantilever oscillation. To avoid crosstalk between topography and recognition images, PicoTREC requires using cantilever with small quality factor (Q≈1). Such low Q, however, results in low force sensitivity, thus a large tip-sample interaction force during imaging, which is destructive to the soft biological samples. In this paper, we propose an innovative method for simultaneous topography imaging and molecular recognition using cantilever with high Q but with minimized crosstalk. For cantilever with large quality factor (Q>10) in liquid, the upper part and lower part of the cantilever´s oscillation becomes symmetric. The topography information that exists in the lower amplitude of the deflection signal also exists in the upper amplitude. Because of the symmetry of the deflection signal, the topography information in the upper amplitude can be cancelled by the lower amplitude. However, the control system will conversely respond to the decrease of the amplitude caused by the molecule binding. So taking use of the lower amplitude, the topography crosstalk in the upper amplitude can be cancelled and the molecule recognition information can be kept. So using this method, high-Q cantilever can be used for simultaneous topography and molecular recognition without crosstalk. The high-Q cantilever will result in high-force sensitivity that will be less destructive to the soft biological samples. This technique will make the simultaneous topography imaging and molecular recognition possible on soft biological sample surfaces, such as the cell membrane. In this paper, the simulat- on is utilized to demonstrate the principle of our proposed method, and then initial experiments are also performed to validate our method.
  • Keywords
    atomic force microscopy; biological techniques; biology computing; biomembranes; cantilevers; cellular biophysics; feature extraction; image recognition; molecular biophysics; tomography; PicoTREC; atomic force microscope; biomolecule recognition; cantilever oscillation; cell membrane; control system; deflection signal symmetry; functionalize tip; high-Q cantilever; high-force sensitivity; large tip-sample interaction force; low force sensitivity; lower amplitude; molecular recognition information; molecule binding; quality factor; recognition image extraction; simultaneous topography imaging; soft biological samples; topography crosstalk; topography image extraction; topography information; upper amplitude; Crosstalk; Force; Image recognition; Microscopy; Sensitivity; Surfaces; AFM; crosstalk; force sensitivity; molecular recognition; quality factor;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nano/Molecular Medicine and Engineering (NANOMED), 2013 IEEE 7th International Conference on
  • Conference_Location
    Phuket
  • Print_ISBN
    978-1-4799-2689-3
  • Type

    conf

  • DOI
    10.1109/NANOMED.2013.6766329
  • Filename
    6766329