Title :
Radio frequency attenuation of saline solutions
Author_Institution :
Amador Corp., St. Paul, MN, USA
Abstract :
Radio frequency (RF) susceptibility tests on pacemakers are frequently performed with the pacemakers in a saline solution. The attenuation provided by three different saline concentrations over the frequency range of 30-800 MHz is investigated. During these investigations it was found that 10 cm of saline provides anywhere from 0 to 19 dB of attenuation depending on the frequency and saline concentration
Keywords :
attenuation measurement; electronic equipment testing; pacemakers; radiofrequency interference; solutions; 30 to 800 MHz; RF susceptibility tests; RFI; UHF; VHF; frequency range; pacemakers; radio frequency attenuation; saline concentrations; saline solutions; Antenna measurements; Attenuation measurement; Frequency measurement; Instruments; Noise measurement; Printed circuits; Protection; Radio frequency; Radiofrequency amplifiers; Wire;
Conference_Titel :
Electromagnetic Compatibility, 1991. Symposium Record., IEEE 1991 International Symposium on
Conference_Location :
Cherry Hill, NJ
Print_ISBN :
0-7803-0158-7
DOI :
10.1109/ISEMC.1991.148209