DocumentCode :
3562262
Title :
Validation of a blood pressure simulator that regenerates oscillometric cuff pressure waveforms
Author :
Dingchang Zheng ; Chengyu Liu ; Amoore, John ; Mieke, Stephan ; Murray, Alan
Author_Institution :
Inst. of Cellular Med., Newcastle Univ., Newcastle upon Tyne, UK
fYear :
2014
Firstpage :
841
Lastpage :
844
Abstract :
Blood pressure (BP) simulators that regenerate oscillometric waveforms provide an alternative for BP device validation. However, their ability to regenerate oscillometric waveforms recorded from unstable conditions has not been fully investigated. This study aimed to provide this information. Manual auscultatory systolic and diastolic blood pressures (SBP and DBP) were measured on 10 healthy subjects under both resting and regular deep breathing conditions. During the manual measurement the oscillometric cuff pressure waveforms were recorded digitally. A specially designed BP simulator was used to regenerate the oscillometric waveforms, which were presented to a clinically validated automatic oscillometric non-invasive BP device to obtain automated BPs from all the 20 waveforms. The SBP and DBP changes induced by deep breathing were finally quantified and compared with the measurement by the automatic device and the manual auscultatory method. Deep breathing decreased both manual and automated SBPs significantly by 5.0 and 6.0 mmHg in comparison with those from the resting condition (both P<;0.01). The corresponding decreases of manual and automated DBPs were 2.6 and 3.3 mmHg (both P<;0.05). The automated BP decrease induced by deep breathing was not significantly diferent from that for manual BP (both P>0.5). Our results demonstrated that the BP simulator can regenerate unstable physiological oscillometric waveforms, confirming that it could be an alternative to clinical trials.
Keywords :
biomedical equipment; blood pressure measurement; pneumodynamics; waveform analysis; waveform generators; BP simulator design; BP waveform; DBP change comparison; DBP change quantification; DBP measurement; SBP change comparison; SBP change quantification; SBP measurement; automated BP; automated DBP reduction; automated SBP reduction; automatic oscillometric noninvasive BP device validation; blood pressure simulator validation; clinical trial; clinical validation; digital cuff pressure waveform recording; manual DBP reduction; manual SBP reduction; manual auscultatory diastolic blood pressure measurement; manual auscultatory systolic blood pressure measurement; oscillometric cuff pressure waveform recording; oscillometric cuff pressure waveform regeneration; oscillometric waveform recording; regular deep breathing condition; resting condition; unstable physiological oscillometric waveform regeneration; unstable waveform recording condition; Abstracts; Clinical trials; Current measurement; Manuals; Physiology;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computing in Cardiology Conference (CinC), 2014
ISSN :
2325-8861
Print_ISBN :
978-1-4799-4346-3
Type :
conf
Filename :
7043174
Link To Document :
بازگشت