• DocumentCode
    3568692
  • Title

    Palm-size low-level neutron sensor for radiation monitoring

  • Author

    Aryaeinejad, R. ; Dardenne, Y.X. ; Cole, J.D. ; Caffrey, A.J.

  • Author_Institution
    Idaho Nat. Eng. Lab., Idaho Falls, ID, USA
  • Volume
    1
  • fYear
    1995
  • Firstpage
    639
  • Abstract
    We have developed a new technique for the detection of neutrons in the presence of a high gamma-ray background. This was accomplished by using two lithium-loaded glass scintillators (6Li/7Li) in combination. The neutron capture reaction in 6Li has a very high Q-value which means better discrimination against gamma-ray background. The 7Li detector, on the other hand, is neutron insensitive and was used to measure separately the gamma contribution in a mixed neutron-gamma field. By subtracting this gamma contribution from the 6Li data, which is sensitive to both gamma rays and neutrons, one can obtain the pure neutron counts
  • Keywords
    glass; lithium; neutron detection; radioactivity measuring apparatus; solid scintillation detectors; Li; high gamma-ray background; lithium-loaded glass scintillators; mixed neutron-gamma field; neutron capture reaction; palm-size low-level neutron sensor; radiation monitoring; Electronic equipment testing; Gamma ray detection; Gamma ray detectors; Gamma rays; Glass; Laboratories; Material storage; Neutrons; Radiation detectors; Radiation monitoring;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
  • Print_ISBN
    0-7803-3180-X
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1995.504288
  • Filename
    504288