DocumentCode
3571056
Title
Monitoring the Change in Resonance Frequency of a Dental Implant in Hardening Dental Plaster Using Microcontroller
Author
Chowdhury, Debasish ; Goswami, Bhaswati ; Ghosh, Ratna
Author_Institution
Dept. of Electron. & Instrum. Eng., Nat. Inst. Sci. & Technol., Berhampur, India
fYear
2014
Firstpage
89
Lastpage
92
Abstract
A low cost instrumentation system was proposed in this paper to determine the stability of dental implants using Resonance Frequency Analysis (RFA) technique. In the present work, an improved system has been described which provides a microcontroller based tracking of the change in resonance frequency (RF) when the dental implant is dipped in a hardening dental plaster mixture. In order to do so, the acquisition of the sensor output using the microcontroller as well as its corresponding display on the LCD has been calibrated and characterized. Thereafter, the performance of the system in tracking the change in RF in in-vitro experiments has been validated.
Keywords
biomedical electronics; biosensors; dentistry; liquid crystal displays; microcontrollers; mixtures; phase detectors; prosthetics; resonance; LCD display; RF change monitoring; RFA technique; dental implants stability; hardening dental plaster mixture; low cost instrumentation system; microcontroller based tracking; phase detector circuit; resonance frequency analysis; resonance frequency change monitoring; sensor output acquisition; Dentistry; Impedance; Implants; Instruments; Microcontrollers; Radio frequency; Resonant frequency; Microcontroller; dental implant; impedance response; instrumentation system; phase detector circuit; resonance frequency;
fLanguage
English
Publisher
ieee
Conference_Titel
Emerging Applications of Information Technology (EAIT), 2014 Fourth International Conference of
Type
conf
DOI
10.1109/EAIT.2014.70