• DocumentCode
    357171
  • Title

    A practical method for calibrating antennas

  • Author

    Prakash, B.N. ; Suresh, K.R.

  • Author_Institution
    EMC Test Centre, Bharat Electron. Ltd., Bangalore, India
  • fYear
    1999
  • fDate
    6-8 Dec. 1999
  • Firstpage
    256
  • Lastpage
    259
  • Abstract
    Calibrated antennas are must for a certified EMC laboratory. A study has been conducted for calibrating antennas in the frequency range 200 MHz-1 GHz, which are needed for 1 meter radiated emission testing-RE02 of MIL-STD-461C. Measurements were conducted to ascertain the antenna factor for conical log spiral antenna by adapting ´Identical Two Antenna Method of calibration´. These measurements were conducted in the frequency range of 200 MHz-1 GHz. The test results indicate the measured antenna factors within +/- 1 dB of the nominal values supplied originally by the manufacturer.
  • Keywords
    antenna testing; calibration; conical antennas; electromagnetic compatibility; military equipment; spiral antennas; 1 m; 200 MHz to 1 GHz; MIL-STD-461C; antenna factor; calibrated antennas; certified EMC laboratory; conical log spiral antenna; identical two antenna method; radiated emission testing; Adaptive arrays; Antenna measurements; Calibration; Electromagnetic compatibility; Electronic equipment testing; Frequency measurement; Gain measurement; Laboratories; Receiving antennas; Transmitting antennas;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electromagnetic Interference and Compatibility '99. Proceedings of the International Conference on
  • Print_ISBN
    81-900652-0-3
  • Type

    conf

  • DOI
    10.1109/ICEMIC.1999.871640
  • Filename
    871640