Abstract :
A precise calibrating system for newly adopted beam position monitors (BPM) of upgraded injection system in NSRL phase II project is described. The fitting error of the electric field is presented. Using the calibrating system and the method which has a system error about ±15μm and the precision of 1μm, we have completed the calibration and checking of the quality to the developed equipment. Due to the beam pipe is fixed vertically and move the antenna to simulate beam orbit deviation, as well as we adopt the RF CW simulating the beam and not the previously 5MHz CW, and the results show fine precision and better reliability
Keywords :
accelerator RF systems; beam handling equipment; calibration; electron accelerators; electron beams; particle beam diagnostics; particle beam injection; storage rings; 5 MHz; BPM calibration; HLS; NSRL phase II project; beam orbit deviation; beam pipe; beam position monitors; electric field; electron beams; error analysis; fine precision; precise calibrating system; radiofrequency simulation; reliability; upgraded injection system; vacuum chamber; Antenna measurements; Electrodes; Error analysis; Extraterrestrial measurements; Fitting; Gratings; High level synthesis; Machine tools; Radio frequency; Signal detection;