    Continuous Blood Pressure Estimation using Pulse Arrival Time and Photoplethysmogram

  • Author

    Chua, C.-P. ; Heneghan, C.

  • Author_Institution
    University College Dublin, Ireland. echua@ee.ucd.ie
  • fYear
    2006
  • Firstpage
    1
  • Lastpage
    5
  • Abstract
    Pulse arrival time and the amplitude of the finger photoplethysmogram were used to track continuous systolic blood pressure over 10-minute intervals. The measures were recorded with a holter-oximeter from a sample of 8 young, healthy human subjects in the supine and standing positions. Results show that, with individual calibration, systolic blood pressure can be estimated with an average error of ±5 mmHg. Using pulse arrival time in combination with the amplitude of the finger photoplethysmogram gave better results than using any one alone.
  • Keywords
    blood pressure; pho-toplethysmogram; pulse arrival time;