• DocumentCode
    3593644
  • Title

    Timing measurements of a micro-channel plate photomultiplier incorporating a novel equal-time anode structure

  • Author

    Byrum, Karen ; Anderson, John ; Drake, Gary ; Ertley, Camden ; Frisch, Henry ; Sanders, Harold ; Tang, Fukun ; Vra, Jerry Va

  • Author_Institution
    Argonne Nat. Lab., Argonne
  • Volume
    1
  • fYear
    2007
  • Firstpage
    80
  • Lastpage
    82
  • Abstract
    We have measured the timing properties on two new 2-inch micro-channel plate photomultiplier tubes from Photonis; one tube with 25 micron pores and the other tube having 10 micron pores. The 10 micron pore tube uses a novel charge- collection scheme at the anode to provide equal arrival time of the signal independent of the hit position of incident light on the face of the tube. We have performed these measurements using a Hamamatsu PLP-10 picosecond laser and a commercial Camac readout electronics system. We present these results, and compare timing properties to a standard Burle/Photonis 85011 micro-channel plate photomultiplier tube having 25 micron pores. We set an upper limit on the resolution of 9.1psec with Npe approximately 50.
  • Keywords
    nuclear electronics; photomultipliers; position sensitive particle detectors; readout electronics; Hamamatsu PLP-10 picosecond laser; MCPPMT; commercial Camac readout electronics system; micro-channel plate photomultiplier tubes; photonis 85011; time anode structure; timing measurements; Anodes; Detectors; Extraterrestrial measurements; Laboratories; Laser beams; Nuclear measurements; Performance evaluation; Photomultipliers; Telephony; Timing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-0922-8
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2007.4436292
  • Filename