• DocumentCode
    3598216
  • Title

    Evaluation of three types of electromagnetic hyperthermia applicator

  • Author

    Johnson, R.H. ; Murfin, J.L. ; Preece, A.W.

  • Author_Institution
    Bristol R. Infirmary, UK
  • fYear
    1988
  • fDate
    4/25/1988 12:00:00 AM
  • Firstpage
    42370
  • Abstract
    The authors compare the theoretical and experimental operation of three types of hyperthermia applicator when heating muscle phantom at 400 and 900 MHz. The three applicators considered are: (i) a waveguide type applicator; (ii) a compact applicator consisting of a half-wavelength resonant patch on high relative permittivity, low loss dielectric substrate backed by a ground plane; (iii) a lightweight inductive applicator
  • Keywords
    biomedical equipment; biothermics; dielectric-loaded waveguides; muscle; physiological models; radiation therapy; radiofrequency heating; 400 MHz; 900 MHz; compact applicator; dielectric loaded waveguide applicator; electromagnetic hyperthermia applicator; half-wavelength resonant patch; lightweight inductive applicator; low loss dielectric substrate; microwave heating; muscle phantom; tumour treatment;
  • fLanguage
    English
  • Publisher
    iet
  • Conference_Titel
    Medical Applications of Microwaves, IEE Colloquium on
  • Type

    conf

  • Filename
    209127