DocumentCode :
3600159
Title :
Use of a precise and portable flow standard to improve the accuracy of medical and scientific laboratory flow measurements
Author :
Conti, J.C. ; Strope, E.R. ; Spence, L.D.
Author_Institution :
Dynatek Dalta Sci. Instrum., Galena, MO, USA
fYear :
1997
Firstpage :
11
Lastpage :
14
Abstract :
The authors have used a portable and precise flow standard to assist in the evaluation of a number of instruments and technologies that are commonly utilized to determine flow and/or velocity of a liquid in both the medical and non-medical laboratories. They have demonstrated the kind of error that one gets from medical ultrasound equipment when random checking of the equipment is performed. They have determined that it is easy for medical ultrasound technologists to incorporate a factor in their peak velocity or flow calculations to convert their qualitative data into reliable quantitative data. The portable flow standard has been used extensively in the authors´ laboratory to allow quick and accurate checking of various electromagnetic, ultrasonic and optical flow evaluation probes. Finally, the authors have utilized this portable flow standard to assist in the initial set-up and debugging of laser-based hydrodynamic systems and have found it indispensable in these particular applications. More work is ongoing evaluating the long-term reliability of this device
Keywords :
biomedical ultrasonics; blood flow measurement; flow measurement; laboratory techniques; laser applications in medicine; measurement standards; device long-term reliability; electromagnetic measurements; flow measurements accuracy improvement; laser-based hydrodynamic systems; medical laboratory flow measurements; medical ultrasound technologists; nonmedical laboratories; optical flow evaluation probes; peak velocity; precise portable flow standard; random equipment checking; reliable quantitative data; scientific laboratory flow measurements; ultrasonic measurements; Biomedical imaging; Fluid flow measurement; Instruments; Laboratories; Laser modes; Modems; Optical design; Probes; Ultrasonic imaging; Velocity measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Engineering Conference, 1997., Proceedings of the 1997 Sixteenth Southern
ISSN :
1086-4105
Print_ISBN :
0-7803-3869-3
Type :
conf
DOI :
10.1109/SBEC.1997.583203
Filename :
583203
Link To Document :
بازگشت