    Parylene-Based Electrochemical-MEMS Force Sensor for Studies of Intracortical Probe Insertion Mechanics

  • Author

    Kim, Brian J. ; Gutierrez, Christian A. ; Meng, Ellis

  • Author_Institution
    Dept. of Biomed. Eng., Univ. of Southern California, Los Angeles, CA, USA
  • Volume
    24
  • Issue
    5
  • fYear
    2015
  • Firstpage
    1534
  • Lastpage
    1544
  • Abstract
    To investigate the mechanical interactions between the implanted cortical multielectrode probes and brain tissue, a Parylene C-based electrochemical-microelectromechanical systems force sensor array was developed. The array consists of seven linearly distributed sensor units arranged along the length of a flexible Parylene C microchannel. The seven sensor units are formed by eight fluidically coupled and adjacent platinum electrode pairs enclosed within the microchannel. Deformation of the top surface of the mechanically compliant microchannel changes the volumetric conduction path between the pairs of sensing electrodes, and therefore, the measured electrochemical impedance, which is proportional to the magnitude of the contacting force. Each sensor unit demonstrated a linear response from 0 to 60 mN with a sensitivity of 0.13 ± 0.01 percentage change in impedance/μN (%/μN; mean ± SE, n = 6). The sensor arrays were mounted onto a ceramic intracortical probe and inserted into the tissue phantoms to verify in situ functionality and assess interfacial probe mechanics. Probe surface force distribution was measured under various insertion speeds and the results indicated that interfacial forces are distributed nonuniformly along the probe shaft length, concentrating within the first 1 mm of the advancing probe tip. Faster insertion speeds were also found to decrease the magnitude of the interfacial forces, suggesting that the tissue strain during cortical implantation may be minimized through appropriate selection of the insertion speed.
  • Keywords
    biological tissues; biomedical transducers; brain; ceramics; electric impedance measurement; electrochemical electrodes; electrochemical sensors; force measurement; force sensors; microsensors; phantoms; polymers; probes; prosthetics; sensor arrays; surface topography measurement; adjacent platinum electrode; brain tissue phantom; ceramic intracortical probe insertion mechanical interaction; distributed sensor; electrochemical impedance measurement; flexible Parylene C microchannel; implanted cortical multielectrode probe; microelectromechanical system; parylene-based electrochemical-MEMS force sensor array; probe surface force distribution measurement; size 1 mm; top surface deformation; volumetric conduction path; Arrays; Electrodes; Impedance; Impedance measurement; Microchannels; Probes; Sensors; Force sensor; Parylene C; electrochemical sensing; electrochemical transduction; electrochemical-microelectro-mechanical systems (EC-MEMS); electrochemical-microelectro-{}mechanical systems (EC-MEMS); impedance-based sensor; liquid impedance; mechanotransduction; mechanotransduction.;
  • fLanguage
    English
  • Journal_Title
    Microelectromechanical Systems, Journal of
  • Publisher
    ieee
  • ISSN
    1057-7157
  • Type

    jour

  • DOI
    10.1109/JMEMS.2015.2420043
  • Filename
    7088554