DocumentCode :
3602783
Title :
A Novel Algorithm for EMG Signal Processing and Muscle Timing Measurement
Author :
Pasinetti, Simone ; Lancini, Matteo ; Bodini, Ileana ; Docchio, Franco
Author_Institution :
Dept. of Mech. & Ind. Eng., Univ. of Brescia, Brescia, Italy
Volume :
64
Issue :
11
fYear :
2015
Firstpage :
2995
Lastpage :
3004
Abstract :
This paper presents a new method for the automated processing of surface electromyography (SEMG) signals, particularly suited for the detection of muscle activation timing. The method has an intermediate level of complexity between simpler (but less performing) and more complex (but in general slower) methods, and is successfully used in the development of biomedical devices for rehabilitation carried out by our group. The method proposed here is based on a statistical approach for threshold computation that is implemented without the need of maximum voluntary contraction or relaxed state, usually required to overcome the difficulty in obtaining the threshold value. The method is compared with 10 popular automated standard methods using different types of simulated signals that approximate the behavior of real SEMG signals. Both the number of activations detected and the onset time measured are analyzed. The algorithm is then applied to real SEMG signals acquired from healthy subjects. The results are finally compared with the literature values. The results show that the proposed algorithm is the best performing method when both the number of activations and the activation timing are considered. In real applications, the algorithm gives the results compatible with the well-agreed literature data.
Keywords :
biomedical equipment; electromyography; medical signal processing; muscle; neurophysiology; statistical analysis; SEMG signal processing; automated processing; biomedical devices; maximum voluntary contraction; muscle activation timing; muscle timing measurement; popular automated standard methods; rehabilitation; signal simulation; statistical approach; surface electromyography signals; Electromyography; Muscles; Signal processing algorithms; Signal to noise ratio; Standards; Timing; Activation analysis; electromyography; measurement; muscles; signal processing; timing; timing.;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2015.2434097
Filename :
7117418
Link To Document :