DocumentCode :
3607179
Title :
Quantitative Verification of Dynamic Wedge Dose Distribution Using a 2D Ionization Chamber Array
Author :
Sahnoun, Tarek ; Farhat, Leila ; Mtibaa, Anis ; Besbes, Mounir ; Daoud, Jamel
Author_Institution :
Dept. of Radiotherapy, Habib Bourguiba Univ. Hosp., Sfax, Tunisia
Volume :
14
Issue :
7
fYear :
2015
Firstpage :
722
Lastpage :
726
Abstract :
The accuracy of two calculation algorithms of the Eclipse 8.9 treatment planning system (TPS)-the anisotropic analytic algorithm (AAA) and pencil-beam convolution (PBC)-in modeling the enhanced dynamic wedge (EDW) was investigated. Measurements were carried out for 6 and 18 MV photon beams using a 2D ionization chamber array. Accuracy of the TPS was evaluated using a gamma index analysis with the following acceptance criteria for dose differences (DD) and distance to agreement (DTA): 3%/3 mm and 2%/2 mm. The TPS models the dose distribution accurately except for 20×20 cm2 field size, 60 ° and 45 ° wedge angles using PBC at 6 MV photon energy. For these latter fields, the pass rate and the mean value of gamma were less than 90% and more than 0.5, respectively at the (3%/3 mm) acceptance criteria. In addition, an accuracy level of (2%/2 mm) was achieved using AAA with better agreement for 18 MV photon energy.
Keywords :
convolution; dosimetry; ionisation chambers; 2D ionization chamber array; Eclipse 8.9 treatment planning system; anisotropic analytic algorithm; distance to agreement; dose differences; dynamic wedge dose distribution; gamma index analysis; pencil beam convolution; photon beams; Accuracy; Arrays; Collimators; Heuristic algorithms; Ionization chambers; Photonics; Planning; 2D ionization chamber array; Calculation algorithms; dose distribution comparison; enhanced dynamic wedge (EDW); gamma index;
fLanguage :
English
Journal_Title :
NanoBioscience, IEEE Transactions on
Publisher :
ieee
ISSN :
1536-1241
Type :
jour
DOI :
10.1109/TNB.2015.2466102