Title :
Experimental Evaluation of a SiPM-Based Scintillation Detector for MR-Compatible SPECT Systems
Author :
Busca, Paolo ; Occhipinti, Michele ; Trigilio, Paolo ; Cozzi, Giulia ; Fiorini, Carlo ; Piemonte, Claudio ; Ferri, Alessandro ; Gola, Alberto ; Nagy, Kalman ; Bukki, Tamas ; Rieger, Jan
Author_Institution :
Dipt. di Elettron., Politec. di Milano, Milan, Italy
Abstract :
In the present work we briefly describe the architecture of a photo-detection module, designed in the framework of the INSERT (INtegrated SPECT/MRI for Enhanced Stratification in Radio-chemoTherapy) project, supported by the European Community. We focus on two main elements of the module: the SiPM photo-detector unit and the multi-channel ASIC. These two components have been investigated with dedicated and independent setups to assess preliminary performance of INSERT architecture. In details, we designed a 25.30 mm ×25.85 mm tile, comprising 9 pixels, each one with an 8 mm ×8 mm active area. We developed an Anger camera to characterize the tile coupled to a CsI:Tl scintillator (6 mm thick). We measured an average spatial resolution (FWHM) of 2 mm in the central region of the Field of View and a 15.3% energy resolution using a 57Co source (122 keV), when the tile is cooled down to 0 ° C to reduce the impact of the dark count rate. Furthermore, we developed ANGUS, a 36-channels 0.35 μm CMOS technology ASIC designed to cope with input capacitance up to 5 nF, typical of large area SiPM pixels. The spectroscopic capability of single readout channels were evaluated by coupling an 8 mm ×8 mm pixel with a cylindrical CsI:Tl scintillator (8 mm diameter, 10 mm thickness). Energy resolution at room temperature provided values between 13% and 13.5% (FWHM) at the 122 keV line for the nine pixels.
Keywords :
application specific integrated circuits; capacitance; nuclear electronics; photomultipliers; readout electronics; scintillation counters; semiconductor counters; Anger camera; CMOS technology; European Community; INSERT architecture; Integrated SPECT/MRI for Enhanced Stratification in Radiochemotherapy project; MR-compatible SPECT systems; SiPM photo-detector unit; SiPM pixels; SiPM-based scintillation detector; cylindrical CsI:Tl scintillator; dark count rate; energy resolution; field of view; input capacitance; multichannel ASIC; photodetection module architecture; single readout channels; spatial resolution; spectroscopic capability; Application specific integrated circuits; Cameras; Detectors; Energy resolution; Magnetic resonance imaging; Single photon emission computed tomography; Spatial resolution; CsI:Tl; SPECT/MRI; SiPM;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.2015.2481184