• DocumentCode
    3608219
  • Title

    A nonlinear lumped model for ultrasound systems using CMUT arrays

  • Author

    Satir, Sarp ; Degertekin, F. Levent

  • Author_Institution
    Dept. of Electr. Eng., Georgia Inst. of Technol., Atlanta, GA, USA
  • Volume
    62
  • Issue
    10
  • fYear
    2015
  • fDate
    10/1/2015 12:00:00 AM
  • Firstpage
    1865
  • Lastpage
    1879
  • Abstract
    We present a nonlinear lumped model that predicts the electrical input-output behavior of an ultrasonic system using CMUTs with arbitrary array/membrane/electrode geometry in different transmit-receive configurations and drive signals. The receive-only operation, where the electrical output signal of the CMUT array in response to incident pressure field is calculated, is included by modifying the boundary element-based vibroacoustic formulation for a CMUT array in rigid baffle. Along with the accurate large signal transmit model, this formulation covers pitch-catch and pulse-echo operation when transmit and receive signals can be separated in time. In cases when this separation is not valid, such as CMUTs used in continuous wave transmit-receive mode, pulse-echo mode with a nearby hard or soft wall or in a bounded space such as in a microfluidic channel, an efficient formulation based on the method of images is used. Some of these particular applications and the overall modeling approach have been validated through comparison with finite element analysis on specific examples including CMUTs with multiple electrodes. To further demonstrate the capability of the model for imaging applications, the two-way response of a partial dual-ring intravascular ultrasound array is simulated using a parallel computing cluster, where the output currents of individual array elements are calculated for given input pulse and compared with experimental results. With its versatility, the presented model can be a useful tool for rapid iterative CMUT-based system design and simulation for a broad range of ultrasonic applications.
  • Keywords
    electrodes; finite element analysis; nonlinear network analysis; ultrasonic transducer arrays; CMUT arrays; arbitrary array/membrane/electrode geometry; boundary element-based vibroacoustic formulation; continuous wave transmit-receive mode; electrical input-output behavior; finite element analysis; intravascular ultrasound array; microfluidic channel; nonlinear lumped model; parallel computing; pulse-echo mode; ultrasonic system; ultrasound systems; Acoustics; Computational modeling; Electrodes; Imaging; Integrated circuit modeling; Mathematical model; Reflection;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2015.007145
  • Filename
    7296775