DocumentCode :
3613066
Title :
A Circadian and Cardiac Intraocular Pressure Sensor for Smart Implantable Lens
Author :
Donida, Achille ; Di Dato, Giuseppe ; Cunzolo, Paolo ; Sala, Marco ; Piffaretti, Filippo ; Orsatti, Paolo ; Barrettino, Diego
Author_Institution :
Department of Technology and Innovation (DTI), University of Applied Sciences of Southern Switzerland, Manno, Switzerland
Volume :
9
Issue :
6
fYear :
2015
Firstpage :
777
Lastpage :
789
Abstract :
This paper presents a new system to measure the Intraocular Pressure (IOP) with very high accuracy (0.036 mbar) used for monitoring glaucoma. The system not only monitors the daily variation of the IOP (circadian IOP), but also allows to perform an spectral analysis of the pressure signal generated by the heartbeat (cardiac IOP). The system comprises a piezoresistive pressure sensor, an application-specific integrated circuit (ASIC) to read out the sensor data and an external reader installed on customized glasses. The ASIC readout electronics combines chopping modulation with correlated double sampling (CDS) in order to eliminate both the amplifier offset and the chopper ripple at the sampling frequency. In addition, programmable current sources are used to compensate for the atmospheric pressure ( 800–1200 mbar ) and the circadian component ( \\pm {7}~{\\rm m\\bar} ) thus allowing to read out the very weak cardiac signals ( \\pm {1.6}~{\\rm m\\bar} ) with a maximum accuracy of 0.036 mbar.
Keywords :
Application specific integrated circuits; Biomarkers; Biomedical measurement; Biomedical telemetry; Lenses; Piezoresistance; Pressure sensors; Biomedical telemetry; glaucoma; implantable lens; intraocular pressure;
fLanguage :
English
Journal_Title :
Biomedical Circuits and Systems, IEEE Transactions on
Publisher :
ieee
ISSN :
1932-4545
Type :
jour
DOI :
10.1109/TBCAS.2015.2501320
Filename :
7384766
Link To Document :
بازگشت