Comparative study of scintillators for PET/CT detectors
Author
A. Nassalski;M. Kapusta;T. Batsch;D. Wolski;D. Mockel;W. Enghardt;M. Moszynski
Author_Institution
Soltan Inst. for Nucl. Studies, Warsaw, Poland
Volume
5
fYear
2005
fDate
6/27/1905 12:00:00 AM
Firstpage
2823
Lastpage
2829
Abstract
A growing interest in the development of dual modality PET/CT scanners prompts the comparative study of numerous scintillators to select the best one, which could be used simultaneously in PET detectors working in the pulsing mode and in the CT detectors working in the current mode. In the comparative measurements, done in the same experimental conditions, various samples of BGO, GSO, GSO:Ce,Zr, LGSO, LSO, LYSO, MLS, LaCl3, LaBr3 and CWO scintillators were tested. The measurements covered a determination of the light output, energy resolution, non-proportionality of the light yield, decay times of the light pulses and for the selected crystals their time resolution for 511 keV annihilation quanta. Moreover, a comparative study of afterglow, induced by 60 keV gamma-rays from a strong 241Am source (13.9 GBq), was done in the second range
Keywords
"Positron emission tomography","Detectors","Pulse measurements","Energy resolution","Computed tomography","Multilevel systems","Testing","Energy measurement","Time measurement","Crystals"