Title :
The design of narrow-band water loaded ultrasound transmitters
Author :
B. Ivancevic;A. Petosic;K. Jambrosic
Author_Institution :
Faculty of Electrical Engineering and Computing, Zagreb, Croatia, e-mail: bojan.ivancevic@fer.hr
fDate :
6/27/1905 12:00:00 AM
Abstract :
The performance of thin-disk transducers made of PZT-5A ceramic using acoustic matching and air-backing is described using simplified 1-D model and experimental results. Several ultrasound transducers have been designed for studying influence of high power ultrasound energy on tissue. PZT-5A ceramic has been used as active element with thickness approximately lambda/2, depending on working frequency (500 kHz, 1 MHz, 2 MHz) and with three different diameter to thickness ratio (D/t) for obtaining different directivity patterns in radiated medium. Theoretical input impedance, conductance and bandwidth of transducer systems has been compared with experimental results. Theoretical results have been derived using Mason´s equivalent circuit
Keywords :
"Narrowband","Ultrasonic imaging","Transmitters","Acoustic transducers","Ultrasonic transducers","Ceramics","Frequency","Impedance","Bandwidth","Equivalent circuits"
Conference_Titel :
Applied Electromagnetics and Communications, 2005. ICECom 2005. 18th International Conference on
Print_ISBN :
953-6037-44-0
DOI :
10.1109/ICECOM.2005.205019