• DocumentCode
    3629291
  • Title

    Analysis of CR dose reduction in pediatric patients, based on computer-simulated noise addition

  • Author

    B. Juste;J.I. Villaescusa;R. Tortosa;R. Miro;G. Verdu

  • Author_Institution
    Chemical and Nuclear Engineering Department at the Polytechnic University of Valencia, Cam? de Vera s/n 46022 Spain
  • fYear
    2008
  • Firstpage
    2697
  • Lastpage
    2700
  • Abstract
    This paper validates a technique to add statistical noise to a Computed Radiography (CR) in order to simulate accurately how the same image would appear if taken at a reduced tube current. To that, a noise addition software has been developed to create lower dose CR using existing pediatric radiographies based on the selection of lower X-ray tube current. The effect of different milliAmpere-seconds (mAs) setting on image quality has been evaluated using the CDMAM 3.4 phantom and the obtained results show good agreements between the simulated and real images in terms of noise measurement. The new CR images allow medical researchers to study how lower dose affects the patient diagnosis.
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Print_ISBN
    978-1-4244-1814-5
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/IEMBS.2008.4649758
  • Filename
    4649758