DocumentCode :
3629945
Title :
Ultratrace measurement using micro-droplet with gas-flow assistance in laser-induced breakdown spectroscopy
Author :
Satoshi Ikezawa;Muneaki Wakamatsu;Yury Zimin;Joanna Pawlat;Toshitsugu Ueda
Author_Institution :
Graduate School of IPS, Waseda University, 2-7-S257, Hibikino, Wakamatsu-ku, Kitakyushu-shi, Fukuoka-ken/JAPAN 808-0135
fYear :
2008
Firstpage :
977
Lastpage :
980
Abstract :
A new setup for laser-induced breakdown spectroscopy (LIBS) is introduced. This setup was designed for high sensitive quantitative analysis using the ink-jet technology and gas-flow assistance. The setup presented here allowed precise detection of a micro-droplet for guiding it into a laser beam spot area. The micro-droplet ejection system created constant volume of sample liquid and taking advantage of liquid physical state; density of the solution can be controlled accurately. That means the constant amount of media was maintained in a micro-droplet. This sampling system allowed generating small droplet (diameter 30 μm) confining the whole fixed volume of the sample material in the laser beam spot area (minimum beam spot diameter 53.2 μm), under the effects of any ambient gas. In this paper, the stability with practical use and improvement of detection limit for LIBS measurement is reported.
Keywords :
"Gas lasers","Electric breakdown","Spectroscopy","Plasma measurements","Voltage","Laser beams","Plasma materials processing","Lenses","Laser theory","Sampling methods"
Publisher :
ieee
Conference_Titel :
Sensors, 2008 IEEE
ISSN :
1930-0395
Print_ISBN :
978-1-4244-2580-8
Type :
conf
DOI :
10.1109/ICSENS.2008.4716605
Filename :
4716605