Title :
Image-based ECG sampling of IVUS sequences
Author :
Aura Hernandez;David Rotger;Debora Gil
Author_Institution :
Computer Vision Center and Department of Computer Science, Universitat Aut?noma de Barcelona, Bellaterra, Spain
Abstract :
Longitudinal motion artifacts in intravascular ultrasound (IVUS) sequences hinders a properly 3D reconstruction and vessel measurements. Most of current techniques base on the ECG signal to obtain a gated pullback without the longitudinal artifact by using a specific hardware or the ECG signal itself. The potential of IVUS images processing for phase retrieval still remains little explored. In this paper, we present a fast forward image-based algorithm to approach ECG sampling. Inspired on the fact that maximum and minimum lumen areas are related to end-systole and end-diastole, our cardiac phase retrieval is based on the analysis of tissue density of mass along the sequence. The comparison between automatic and manual phase retrieval (0:07 plusmn 0:07 mm. of error) encourages a deep validation contrasting with ECG signals.
Keywords :
"Electrocardiography","Image sampling","Ultrasonic imaging","Image reconstruction","Motion measurement","Ultrasonic variables measurement","Hardware","Image processing","Image retrieval","Signal sampling"
Conference_Titel :
Ultrasonics Symposium, 2008. IUS 2008. IEEE
Print_ISBN :
978-1-4244-2428-3
DOI :
10.1109/ULTSYM.2008.0322