Title : 
Comparison of ´electrophysiologic imaging´ and radionuclide imaging in quantification of controlled cardiac ischemia
         
        
            Author : 
C.J. Penney;L.M. Title;S.E. Iles;J.C. Clements;M.J. Gardner;B.M. Horacek
         
        
            Author_Institution : 
Dept. of Psychol., Dalhousie Univ., Halifax, NS, Canada
         
        
        
        
        
            Abstract : 
The objective of this study was to determine the utility of body-surface potential mapping (BSPM) in detecting, localizing and quantifying the extent/severity of acute myocardial ischemia during coronary occlusion. Percutaneous transluminal coronary angioplasty (PTCA) was used as a model of ´controlled´ acute myocardial ischemia. We compared the ability of ´electrophysiologic imaging´ (BSPM enhanced by inverse calculation of epicardial potentials) and radionuclide imaging to detect and quantify regional myocardial ischemia. All results were calculated from a polar display of the epicardium. The study group included 18 patients with single-vessel coronary artery disease. ´Electrophysiologic imaging´ poses no risk for the patient and in this study correlates with radionuclide imaging as a method of detecting and localizing acute myocardial ischemia during PTCA.
         
        
            Keywords : 
"Radio control","Ischemic pain","Torso","Myocardium","Arteries","Coronary arteriosclerosis","Anatomy","Electrocardiography","Physiology","Biophysics"
         
        
        
            Conference_Titel : 
Computers in Cardiology 1995
         
        
            Print_ISBN : 
0-7803-3053-6
         
        
        
            DOI : 
10.1109/CIC.1995.482576