Title :
An optimized temperature calibration system
Author :
J. DrnovSek;J. Bojkovski;I. Turk
Abstract :
The proposed paper deals with an optimization of a temperature
calibration system in terms of stability, measurement uncertainty and
calibration time. The system consists of two liquid baths for the
temperature range between -80°C to +290°C. An algorithm for
analyzing maximal stability of calibration media was developed and a
dedicated micro controller unit designed. Digital processing of
temperature data was studied in order to optimize accuracy of
measurement versus calibration time and temperature gradient and
differential measurement. The system was analyzed by Spice simulation