Title :
Features of pulse plethysmography and a comparison with impedance plethysmography
Author :
Ante Šantić;Miroslav Šaban
Author_Institution :
Faculty of Electrical Engineering, University of Zagreb, 3 Unska, 41000 Zagreb, Croatia
Abstract :
The features of current pulse plethysmography are given. The skin capacitance and resistance are depending upon pulse duration but not the internal resistance. The internal resistance changes depend of arteries volume changes. A comparison between pulsed plethysmography and sine-wave plethysmography is discussed. The advantages of pulsed plethysmography like much lower current load through the tissue, higher sensitivity and less noise in the recorded signal versus impedance sine-wave plethysmography is pointed out.
Keywords :
Pulse measurements
Conference_Titel :
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Print_ISBN :
0-7803-0785-2
DOI :
10.1109/IEMBS.1992.5762017