• DocumentCode
    3641415
  • Title

    Time-of-flight measurements with Cherenkov photons produced by 511 keV photons in lead crystals

  • Author

    R. Dolenec;S. Korpar;P. Krizan;R. Pestotnik;A. Stanovnik;R. Verheyden

  • Author_Institution
    Jož
  • fYear
    2010
  • Firstpage
    280
  • Lastpage
    284
  • Abstract
    The resolution of time-of-flight measurements is usually limited by the decay time constant of the scintillating material used. If Cherenkov photons, which are produced promptly, could be used to measure the time-of-flight, the limiting factor would become the photo detector. The excellent time resolution of the microchannel plate photomultiplier (MCP PMT) lead us to consider the possibility of performing such time-of-flight measurements for pairs of annihilation 511 keV photons, which are used in positron emission tomography (PET) for source distribution reconstruction. In this work we present simulation and experimental results, obtained with a simple back-to-back detector, using PbF2 or PWO crystal as Cherenkov radiator and Hamamatsu MCP PMT as a photo detector.
  • Keywords
    "Photonics","Crystals","Surface treatment","Detectors","Photodetectors","Timing"
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4244-9106-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2010.5873765
  • Filename
    5873765