DocumentCode
3641429
Title
Validation of PeneloPET positron range estimations
Author
J. Cal-González;J. L. Herraiz;S. España;M. Desco;J. J. Vaquero;J. M. Udías
Author_Institution
Grupo de Fí
fYear
2010
Firstpage
2396
Lastpage
2399
Abstract
Positron range depends on the materials in which positron propagates and on positron emitter isotope. As positron range limits the spatial resolution of PET images, good quantitative estimates of it should be included in any realistic simulation of PET acquisitions. In this work we compare positron range estimates obtained with PeneloPET to previous available simulations and experimental data. PeneloPET was used to simulate the positron range of 18F, 11C, 13N, 15O, 68Ga and 82Rb in the following tissues: cortical bone, soft bone, skin, muscle, brain, water, adipose tissue and lung. The 3D and 1D annihilation Point Spread Functions (aPSF) were calculated for each isotope-material combination. We have studied with more detail the 3D aPSF (radial distributions) and the cumulative fraction of annihilation events. These aPSF distributions were also studied for non-uniform media. Results obtained were consistent with other results previously reported in the literature as well as with experimentally measured data.
Keywords
"Positrons","Three dimensional displays","Isotopes","Positron emission tomography","Media","Fitting"
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
ISSN
1082-3654
Print_ISBN
978-1-4244-9106-3
Type
conf
DOI
10.1109/NSSMIC.2010.5874216
Filename
5874216
Link To Document