Title :
Linearity and energy resolution of compton electrons in CZT measured using the wide angle compton coincidence technique
Author :
Marek Szawlowski;Maciej Kapusta;Lukasz Swiderski;Radoslaw Marcinkowski;Marek Moszyński;Tomasz Szczęśniak;Martyna Grodzicka;Dariusz Wolski;Anna Celler
Author_Institution :
Soltan Institute for Nuclear Studies, PL 05-400 Ś
Abstract :
The Wide-Angle Compton Coincidence (WACC) technique was adapted to measure the electrons produced in cadmium zinc telluride (CZT) detector by Compton scattered gamma rays of 662 keV energy emitted by a 137Cs source. The measurements were carried out with a 4 × 4 pixel anode array build on a 5 mm thick CZT crystal with 1.1 mm pitch. A High Purity Germanium (HPGe) detector operating in coincidence mode measured energy of the scattered photons. The coincidence pulse height data from the CZT and HPGe detectors were recorded in a list mode. The CZT electron spectra were determined by subtracting the energy of the scattered photons from the incident photon energy. The scattering angle was determined by selecting a narrow energy window in the collected HPGe spectra. The CZT electron linearity and resolution were measured in the energy range of 100 keV-300 keV corresponding to the scattering angle range of 31°-66°.
Keywords :
"Energy resolution","Detectors","Pixel","Photonics","Energy measurement","Linearity","Logic gates"
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
Print_ISBN :
978-1-4244-9106-3
DOI :
10.1109/NSSMIC.2010.5874540