Title :
Statistical image reconstruction methods for simultaneous emission/transmission PET scans
Author :
H. Erdogan;J.A. Fessler
Author_Institution :
Dept. of Electr. Eng. & Comput. Sci., Michigan Univ., Ann Arbor, MI, USA
Abstract :
Transmission scans are necessary for estimating the attenuation correction factors (ACFs) to yield quantitatively accurate PET emission images. To reduce the total scan time, post-injection transmission scans have been proposed in which one can simultaneously acquire emission and transmission data using rod sources and sinogram windowing. However, since the post-injection transmission scans are corrupted by emission coincidences, accurate correction for attenuation becomes more challenging. Conventional methods (emission subtraction) for ACF computation from post-injection scans are suboptimal and require relatively long scan times. The authors introduce statistical methods based on penalized-likelihood objectives to compute ACFs and then use them to reconstruct lower noise PET emission images from simultaneous transmission/emission scans. Simulations show the efficacy of the proposed methods. These methods improve image quality and SNR of the estimates as compared to conventional methods.
Keywords :
"Image reconstruction","Positron emission tomography","Attenuation","Thorax","Vectors","Statistical analysis","Image quality","Yield estimation","Computational modeling","Signal to noise ratio"
Conference_Titel :
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
Print_ISBN :
0-7803-3534-1
DOI :
10.1109/NSSMIC.1996.587926