    Gamma-ray and electron response in doped alkali halide scintillators

  • Author

    L. Swiderski;M. Moszyński;W. Czarnacki;M. Szawłowski;T. Szczęśniak;G. Pausch;C. Plettner;K. Roemer;P. Schotanus

  • Author_Institution
    National Centre for Nuclear Research (former Soltan Institute for Nuclear Studies), Otwock-Swierk, PL-05-400 Poland
  • fYear
    2011
  • Firstpage
    982
  • Lastpage
    986
  • Abstract
    The wide angle Compton coincidence technique was used to study non-proportionality of light yield and intrinsic resolution of doped alkali halides. The tested scintillators were 1 inch × 1 inch cylindrical crystals of NaI:Tl, CsI:Na and CsI:Tl. The experimental setup comprised a High Purity Germanium detector and the tested samples coupled to a photomultiplier. The results obtained for Compton electrons are compared with the response of the tested scintillators to γ-rays. In this study we pointed out the correlation between non-proportionality at low energy range and intrinsic energy resolution measured for Compton electrons and γ-rays. Substantial difference in intrinsic resolution of doped NaI and CsI crystals was observed.
  • Keywords
    "Energy measurement","Energy resolution","Electronic mail"