DocumentCode :
3645913
Title :
2×2 MPPC arrays in gamma spectrometry with CsI(Tl), LSO:Ce(Ca), LaBr3, BGO
Author :
M. Grodzicka;M. Moszyński;T. Szczęśniak;M. Szawłowski;D. Wolski;J. Baszak
Author_Institution :
National Centre for Nuclear Research, Andrzeja Soltana 7, PL 05-400 Otwock - Swierk, Poland
fYear :
2011
Firstpage :
1683
Lastpage :
1692
Abstract :
The aim of this work is characterization of MPPC arrays with active area of 6×6 mm in gamma-ray spectrometry with four scintillators CsI(Tl), LSO:Ce(Ca), LaBr3 and BGO. The chosen scintillators have significantly different decay times of a scintillation pulse and different wavelength of emission peaks. It allowed study of the influence of MPPC effective dead time on a device response and calculation of a Photon Detection Efficiency (PDE) of MPPC. The measurements covered : a selection of the optimum operating voltage, a verification of the linearity range for a given decay time of the crystals, a verification of the excess noise factor for the optimal voltage and calculation of the PDE of MPPC. Measurements of energy resolution and non-proportionality obtained with the MPPC arrays and CsI(Tl) or BGO were compared to the similar measurements performed with XP2020Q PMT.
Keywords :
"Temperature measurement","Crystals","Energy measurement","Laser modes","Voltage control","Spectroscopy","Logic gates"
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
ISSN :
1082-3654
Print_ISBN :
978-1-4673-0118-3
Type :
conf
DOI :
10.1109/NSSMIC.2011.6154661
Filename :
6154661
Link To Document :
بازگشت