DocumentCode :
3646143
Title :
SAR levels in microwave breast imaging: 3-D safety assessment with plane-wave illumination
Author :
Adam Santorelli;Michael Vander Schueren;Milica Popovic
Author_Institution :
Department of Electrical and Computer Engineering, McGill University, Montreal, Canada
fYear :
2011
Firstpage :
481
Lastpage :
484
Abstract :
In this work, we present results of simulation analysis to assess the safety of the microwave-based breast cancer detection techniques. In our numerical experiments, a detailed, three-dimensional anatomical model, derived from magnetic resonance images, has been exposed to plane waves from different angles and at varying frequencies. These simulations result in profiles of specific absorption rate (SAR), a quantity that is used to measure the compliance with safety standards. Our analysis implies that the power levels needed for the functioning of microwave breast cancer detection systems remains safe in terms of acceptable SAR levels.
Keywords :
"Breast","Microwave imaging","Microwave theory and techniques","Numerical models","Safety","Electromagnetic heating"