DocumentCode
3647220
Title
Comparison of methods of piezoelectric coefficient measurement
Author
Jirí Fialka;Petr Beneš
Author_Institution
Department of Control and Instrumentation, Brno University of Technology, Czech Republic
fYear
2012
fDate
5/1/2012 12:00:00 AM
Firstpage
37
Lastpage
42
Abstract
The paper compares the main methods used for measuring of the piezoelectric material constants. It outlines the principle of three measuring methods most commonly used today, i.e. the frequency method, the laser interferometry method and the quasi-static method. These methods have been practically applied to piezoelectric ceramic samples. The paper describes the production of the piezoelectric ceramic samples of defined sizes in accordance with the current regulations. An NCE51 production code soft ceramic was used in the experiments. A piezoelectric charge coefficient was measured. The final values of the piezoelectric charge coefficient obtained through all the methods were compared to the catalogue values of the piezoelectric ceramic. All three methods can be described as appropriate; compared with the frequency method, the laser interferometry and quasi-static methods are rather time-consuming and more demanding with respect to preparation of the measurement experiment. The frequency method provides results within a smaller value range.
Keywords
"Charge measurement","Voltage measurement","Frequency measurement","Ceramics","Measurement by laser beam","Force"
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference (I2MTC), 2012 IEEE International
ISSN
1091-5281
Print_ISBN
978-1-4577-1773-4
Type
conf
DOI
10.1109/I2MTC.2012.6229293
Filename
6229293
Link To Document