Abstract
Based on the analysis of possible mechanical misalignments of a CT (computer tomography) unit, CT image distortions are modeled. By extending the model with respect to the individual image acquisition process, the influence of patient´s weight on CT table velocity is included as well as gentry and table motion tilts. The effect of non-linear bending of the CT table due to patient´s weight is investigated in detail. A method to determine the magnitude of geometric image distortions is proposed. It can be interpreted as an estimation of parameters of a non-linear model in three dimensional image space. The validity and applicability of the correction method, e.g. for ´frameless´ image-guided neurosurgery, is discussed. Applying the method to CT images of different patients, image distortions up to 2 mm have been detected.
Keywords
"Computed tomography","Nonlinear distortion","Neurosurgery","Neoplasms","Engineering in Medicine and Biology Society","Solid modeling","Image enhancement","Electronic mail","Image analysis","Biology computing"