• DocumentCode
    3652313
  • Title

    Temperature dependence of metrological characteristics of quantum Hall effect samples

  • Author

    J. Bohacek;P. Svoboda;P. Vasek

  • Author_Institution
    Dept. of Meas., Czech Tech. Univ., Prague, Czech Republic
  • Volume
    2
  • fYear
    1998
  • Firstpage
    1254
  • Abstract
    Influence of temperature on frequency performance of an AlGaAs-GaAs heterostructure fabricated by Laboratoires d´Electronique Philips has been investigated. The structure was placed into a /sup 4/He bath and significant improvement of its frequency performance has been observed when the operational temperature was reduced from 4.2 K to 1.6 K. For T=1.6 K and I=30 /spl mu/A, an approximately linear frequency dependence of the quantized Hall resistance R/sub H/(2) was found, its slope being 0.11 ppm/kHz.
  • Keywords
    "Temperature dependence","Hall effect","Frequency measurement","Electrical resistance measurement","Calibration","Resistors","Superconducting magnets","Bridges","Helium","Magnetic flux"
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1998. IMTC/98. Conference Proceedings. IEEE
  • ISSN
    1091-5281
  • Print_ISBN
    0-7803-4797-8
  • Type

    conf

  • DOI
    10.1109/IMTC.1998.676925
  • Filename
    676925