Image quantification in high-resolution PET assessed with a new anthropomorphic brain phantom
Author :
Jarkko Johansson;Jarmo Teuho;Jani Linden;Uygar Tuna;Tuula Tolvanen;Virva Saunavaara;Mika Teras
Author_Institution :
Turku PET Centre, Turku University Central Hospital and University of Turku, Kiinamyllynkatu 4-8, 20540, Finland
fYear :
2013
Firstpage :
1
Lastpage :
7
Abstract :
Choice of the PET scanner and image reconstruction parameters have significant impact in quantitative positron-emission tomography (PET). Hoffman phantom is probably the most widely used test object for assessing this impact in brain PET studies. In high-resolution PET, however, it´s usability is questionable due to lesser partial-volume effect. Futhermore, Hoffman phantom is cylindrical and does not offer realistic attenuation effect for the skull. In the current work we used a novel brain phantom that was produced using a 3D-printer, and provides realistic head contour and skull attenuation effect. We scanned the phantom with latest generation whole-body PET/MR (Philips Ingenuity TF) and PET/CT (GE Discovery 690) scanners and in a brain dedicated high-resolution scanner (Siemens HRRT) to evaluate it´s usability for intra- and inter-scanner comparisons with regard to PET brain imaging.
Keywords :
"Image reconstruction","Positron emission tomography","Phantoms","Convergence","Computed tomography","Image resolution","Attenuation"
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE