• DocumentCode
    3658979
  • Title

    A hysteresis current control based shunt current compensation scheme for power quality improvement in high power radiology applications

  • Author

    Shridevi R. Angadi; Nanjundaswamy R.;N. R. Srinivas;R. S. Ananda Murthy

  • Author_Institution
    Department of EEE, SJCE, Mysore, India
  • fYear
    2015
  • Firstpage
    53
  • Lastpage
    58
  • Abstract
    In this paper, a shunt active power filter scheme is proposed for a three phase system feeding a non-linear load. The system represents the HV generator used in high power radiology equipment. Hysteresis current control is used, which controls the operation such that a sinusoidal current is drawn from the source. The proposed method also reduces the THD, thus improving the power quality. The compensating currents are injected in phase opposition at the point of common contact to cancel out the harmonic currents in ac lines. The Hysteresis current control is used because it is simple and easy to implement. The simulation is carried out using MATLAB-SIMULINK software, whose results are displayed.
  • Keywords
    "Switches","Harmonic analysis","Power harmonic filters","Reactive power","Hysteresis","Current control","Insulated gate bipolar transistors"
  • Publisher
    ieee
  • Conference_Titel
    Power and Advanced Control Engineering (ICPACE), 2015 International Conference on
  • Type

    conf

  • DOI
    10.1109/ICPACE.2015.7274916
  • Filename
    7274916