• DocumentCode
    3661578
  • Title

    A New Implementation of A-Mode Ultrasound Pulser-Receiver System

  • Author

    Huihui Li;Yongjin Zhou;Lei Wang;Xiaqing Wen

  • Author_Institution
    Shenzhen Inst. of Adv. Technol., Shenzhen, China
  • fYear
    2014
  • Firstpage
    187
  • Lastpage
    190
  • Abstract
    Ultrasound represents a very compact, low cost method that is non-invasive and does not involve ionizing radiation. A new A-mode ultrasound pulser-receiver system, which digital control module is based on FPGA chip, has been designed and implemented. The pulser-receiver system consists of three parts: digital signal processing module, high frequency pulser and receiver module. The pulser was programmed to generate unipolar pulse trains with 1 cycle, at frequency of 10 MHz, repetition frequency of about 1 KHz, and amplitude of 80 V to drive a 10MHz piezoelectric transducer. Clear ultrasonic pulse echo has been measured in the receiver. Simulations of key components and measured signal of pulser-receiver system have been shown. These results indicate that the system implementation has reached a good performance.
  • Keywords
    "Ultrasonic imaging","Field programmable gate arrays","Receivers","Transmitters","MOSFET","Transducers","Ultrasonic variables measurement"
  • Publisher
    ieee
  • Conference_Titel
    Intelligent Systems, Modelling and Simulation (ISMS), 2014 5th International Conference on
  • ISSN