DocumentCode :
3662724
Title :
A single-ended chopper-stabilized ISFET amplifier for continuous pH measurement applications
Author :
Mohammadreza Asgari; Kye-shin Lee
Author_Institution :
Dept. of Electrical and Computer Engineering, The University of Akron, OH, USA
fYear :
2015
Firstpage :
1
Lastpage :
4
Abstract :
This paper introduces a novel low-power Ion-Sensitive Field Effect Transistor (ISFET) readout configuration compatible with standard CMOS technology that utilizes chopping technique to reduce the 1/f noise and the offset of the readout circuitry. In addition, the proposed scheme tackles the well-known non-idealities of the ISFET such as the trapped charge and the long-term drift by using a resetting switch, while suppressing the imperfections of the resetting switch with the proposed chopping amplifier. The readout circuitry can be applied to both continuous time or difference pH measurements. Simulation results using standard 0.35μm CMOS technology shows 20 dB reduction in the output noise while the ISFET readout circuit including the integrator consumes 950nW.
Keywords :
"CMOS integrated circuits","MOSFET","Switches","Noise","Electrodes","Sensors"
Publisher :
ieee
Conference_Titel :
Circuits and Systems (MWSCAS), 2015 IEEE 58th International Midwest Symposium on
Type :
conf
DOI :
10.1109/MWSCAS.2015.7282183
Filename :
7282183