• DocumentCode
    3664381
  • Title

    Analysis of non-insertable device under test measurement error with vector network analyzer

  • Author

    Haiyang Zhang;Jingkun Liu;Jian Jiang

  • Author_Institution
    The 41st Institute of China Electronics Technology Group Corporation, No. 98, Xiangjiang Road, Economic &
  • fYear
    2015
  • fDate
    5/1/2015 12:00:00 AM
  • Firstpage
    276
  • Lastpage
    280
  • Abstract
    The paper introduces the 12 error terms calibration model algorithm of vector network analyzer(VNA) based on the signal flow graph theory and the Mason formula, and proposes the full 2-port calibration method. The influence of the error caused by measuring non-insertable DUTs with the VNA to the measuring results is analyzed and a corresponding compensation algorithm is given. Meanwhile, this paper discusses how the transmission measurement error is affected respectively by hardware performance of the VNA, the reflection coefficient and transmission coefficient of the through calibration kit and the S parameters of the DUT. Finally, an algorithm which can eliminate the unknown through is presented.
  • Keywords
    "Calibration","Ports (Computers)","Switches","Measurement uncertainty","Reflection coefficient","Loss measurement","Measurement errors"
  • Publisher
    ieee
  • Conference_Titel
    Electronics Information and Emergency Communication (ICEIEC), 2015 5th International Conference on
  • Print_ISBN
    978-1-4799-7283-8
  • Type

    conf

  • DOI
    10.1109/ICEIEC.2015.7284538
  • Filename
    7284538