Title : 
Characterization of CMOS electrochemical oxygen sensor for biomedicai applications
         
        
            Author : 
Colin Chue;Chemmanda John Leo;Wai Pan Chan;Maria Ramona B Damalerio;Ming-Yuan Cheng;Jia Hao Cheong;Yuan Gao
         
        
            Author_Institution : 
Institute of Microelectronics, A∗
         
        
        
            fDate : 
6/1/2015 12:00:00 AM
         
        
        
        
            Abstract : 
This paper presents the characterization procedures of a CMOS electrochemical oxygen sensor and its readout circuits in application specific integrated circuit (ASIC) format. As the sensor needs to be tested inside deionized water, a master controller Printed Circuit Board (PCB) is designed to provide power and control signals to test the ASIC. The transimpedance (TIA) amplifier gain, voltage amplifier gain, sensor array sensitivity and 8-bit Digital-to-Analog Converter (DAC) trimming voltage are configured through the master controller. The oxygen sensor ASIC is mounted on another PCB with chip-on-board (COB) packaging. Biocompatible hydrogel Nafion is coated to cover the entire sensor electrodes for oxygen transportation. The measured sensor sensitivity is 0.78 mV/mmHg and a transient output voltage change of 50 mV is observed when the deionized water is changed from air saturation to nitrogen saturation.
         
        
            Keywords : 
"Conferences","Electron devices","Solid state circuits"
         
        
        
            Conference_Titel : 
Electron Devices and Solid-State Circuits (EDSSC), 2015 IEEE International Conference on
         
        
            Print_ISBN : 
978-1-4799-8362-9
         
        
        
            DOI : 
10.1109/EDSSC.2015.7285116