DocumentCode
3665028
Title
Determination of blackbody cavity reflectance at 633nm by integrating sphere reflectometer
Author
Wang Jing-hui;Yuan Zun-dong;Duan Yuan-yuan;Liu Zeng-lin;Lu Xiao-feng
Author_Institution
Key Laboratory of Thermal Science and Power Engineering of Ministry of Education, Tsinghua University, Beijing, China
fYear
2015
fDate
7/1/2015 12:00:00 AM
Firstpage
120
Lastpage
123
Abstract
The blackbody cavity emissivity is a primary parameter of blackbody source. The measurement of blackbody cavity emissivity could verify the design value of cavity emissivity. A determination of graphite blackbody cavity emissivity by reflectance measurement is presented. The measuring equipment include a He-Ne laser, laser power stabilizer, integrating sphere, Si photoelectric detector, gray reflectance standard sample and Keithley 6485 picoammeter. The preliminary measurement value of graphite blackbody cavity emissivity is 0.99983351.
Keywords
"Cavity resonators","Reflectivity","Detectors","Standards","Graphite","Current measurement","Power measurement"