DocumentCode :
3671958
Title :
Fabrication of 60 channel microelectrode arrays for future use with cultured neuronal networks
Author :
V. P. Gomes;A. D. Barros;J. B. D. Filho;S. Martinoia;A. Pasquarelli;J. W. Swart
Author_Institution :
School of Electr. &
fYear :
2015
Firstpage :
1
Lastpage :
4
Abstract :
Stimulation and recording of nerve cells is a procedure used for several applications, such as clinical therapies, study of basic neuroscience, and neural prostheses. Microtechnology and advances in material science have allowed to produce more sophisticated devices and with more functions. This paper focuses on the fabrication of planar 60-channel MEAs. The electrical characterization of the noise level from the TiN electrodes showed good sensitivity to noise, compatible with commercial systems. These electrodes received an artificial electrocardiogram signal (ECG) from a function generator and registered the same input signal but with lower amplitude. Finally, both cyclic voltammetry curves of the produced MEA and the commercial MEA exhibited similar shape, but the amplitude of the first one was significantly higher than in MCS-MEA, with an order difference of magnitude.
Keywords :
"Semiconductor device measurement","Monitoring","Biomedical monitoring","Electric variables measurement","Electrodes","Gain measurement"