Title :
Efficient wireless power delivery and reliable telemetry for implants and safety
Author :
Rangarajan Jegadeesan;Yong-Xin Guo;Shih-Cheng Yen;Nitish Vyomesh Thakor
Author_Institution :
Singapore Institute for Neurotechnology, National University of Singapore, Singapore-117456
Abstract :
Wireless power delivery and telemetry are the key aspects of an implantable device that enables long-term operation which is otherwise limited by the factors like battery capacity (in battery operated systems), infections and pain (tethered systems). In this paper, we present our efficient wireless power delivery and reliable telemetry platform that enables a completely implantable muscle stimulator. 100 mW of rectified power is delivered to the implant at an efficiency of over 40% and supports bidirectional telemetry at 2.4 Kbps with a bit error rate lesser than 10-7. The wireless platform is tested chronically in rats. Electrical safety is ensured using heating measurements and they augur well with simulation results.
Keywords :
"Implants","Telemetry","Wireless communication","Muscles","Skin","Power system reliability","Reliability"
Conference_Titel :
RF and Wireless Technologies for Biomedical and Healthcare Applications (IMWS-BIO), 2015 IEEE MTT-S 2015 International Microwave Workshop Series on
DOI :
10.1109/IMWS-BIO.2015.7303757