• DocumentCode
    3675932
  • Title

    A wireless solution for intraoperative monitoring

  • Author

    Jeffrey Mays;Patricia Rampy;Dan Sucato;Steven Sparagana;J.-C. Chiao

  • Author_Institution
    Electrical Engineering, The University of Texas at Arlington, Arlington, TX, USA
  • fYear
    2015
  • Firstpage
    244
  • Lastpage
    245
  • Abstract
    Intraoperative neurophysiological monitoring (IONM) is a useful tool for monitoring the human nervous system during surgery. Current IONM systems use wired electrodes to record neural signals. However, these wires subject IONM waveforms to electromagnetic interference (EMI) from operating room (OR) equipment, which distorts recorded signals. A recording system that can be placed near the electrodes would reduce EMI significantly, and clear obstructions due to wires in the OR. In this work, we developed a wireless module for recording IONM signals, and demonstrated that the wireless system has improved resilience to EMI over traditional wired recording systems.
  • Keywords
    "Wireless communication","Electromagnetic interference","Surgery","Electrodes","Wires","Monitoring","Distortion"
  • Publisher
    ieee
  • Conference_Titel
    RF and Wireless Technologies for Biomedical and Healthcare Applications (IMWS-BIO), 2015 IEEE MTT-S 2015 International Microwave Workshop Series on
  • Type

    conf

  • DOI
    10.1109/IMWS-BIO.2015.7303867
  • Filename
    7303867